Ecotron - Incubation Shaker
Operation
03 May 2018
Page
51
of
90
7.2.3 Fitting the Tray
The table features threaded holes which make it possible to mount
holders and clamps directly on the table. Alternatively, trays with
fixtures can be ordered from the manufacturer and fitted onto the
table. To do so, proceed as follows:
Procedure
1.
Place the tray on the table so that no edge sticks out.
2.
Fix the tray to the table using the supplied Phillips-head
screws. Evenly tighten the Phillips-head screws cross-wise.
ATTENTION
If the tray is not affixed securely to the table, the tray may dam-
age the interior of the table when moving.
– Do not start up the equipment unless the tray has been in-
serted correctly and affixed.
Screw sizes
Tray with pin holders: Philips-head M4x8
Tray with sliding bars: Philips-head M4x8
Tray with «Sticky Stuff» adhesive matting: Philips-head M4x6
7.2.4 Fitting the Intermediate Base
An intermediate base for static incubation is available as an acces-
sory. The maximum load of the intermediate base is 5 kg for point
loads. To prevent longer lasting cultures from drying out, we rec-
ommend placing suitable sealing tape (e.g. parafilm) on the agar
plates or placing the plates in an open plastic bag.