Ecotron - Incubation Shaker
Safety and Responsibility
Page
10
of
90
03 May 2018
To use the equipment for special applications not covered by con-
ventional, intended use, the equipment must be modified and certi-
fied accordingly by the manufacturer.
Any use of the equipment outside of a biological laboratory, i.e. in
any environment in which the conditions required for the safety of
users cannot be met or cannot be met to their full extent, shall also
be deemed misuse.
2.2 Cultivation Vessels to Be Used
Significant forces are applied to cultivation vessels, in particular in
case of large vessels and high speeds. Hence, the cultivation ves-
sels used are particularly significant in relation to user safety.
ATTENTION
Use of unsuitable or defective cultivation vessels can lead to
glass breakage and therefore damage to property.
Approved cultivation vessels
The equipment has been designed for use with the following ves-
sels using the holders designed specifically for them:
Erlenmeyer flasks up to 5,000 mL made of borosilicate glass
(e.g. Schott Duran
®
) or high-grade plastic, such as polycar-
bonate (z. B. Corning
®
) etc.
Fernbach flasks up to 3,000 mL made of borosilicate glass
(e.g. Schott Duran
®
) or high-grade plastic, such as polycar-
bonate (z. B. Corning
®
) etc.
Other vessels with the holders designed for them:
Test tubes
Centrifuge tubes
Microtitre plates
Deep well plates
To avoid the vessels coming out of the clamps at very high speeds,
they might have to be secured using cable ties underneath the
springs or some other suitable measure.