7.7 Behaviour in Case of Interrupted Power Supply
If the power supply to the equipment is interrupted during a running
cultivation process (e.g. by flicking the main switch or in case of a
power failure), all parameter setpoints are stored. If the power sup-
ply is restored, the equipment automatically restarts with the last
stored setpoints.
7.8 Using the Equipment as a Tempering Cover
If you dismantle the shaker, the equipment can be used as a pure
tempering cover. To do so, proceed as follows:
Procedure
1.
Turn off equipment and pull out the plug.
2.
Open the door to the incubation chamber.
3.
Remove the table incl. any mounted tray, drip tray and drip
pan (see chapter 10.4 "Replacing the Drip Tray", page 75).
4.
Remove the deflector plate on the ceiling of the incubation
chamber by loosening the two Phillips-head screws (M4x10).
5.
Loosen the two Phillips-head screws (M4x10) that affix the ca-
ble to the side wall.