Ecotron - Incubation Shaker
Rectifying Faults
03 May 2018
Page
63
of
90
Error during temperature
measurement.
Check the function of the temperature sensor. In
case of a defect, contact your local Infors repre-
sentative.
Operator
Heater fan not
working.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Operator
Alarm message
Meaning
Temperature too high (actual value > se 1.5°C).
Possible cause
Measures
Personnel
The parameter has recently
been changed, and the re-
quired setpoint has not been
reached yet.
No action required. The alarm message will dis-
appear as soon as the setpoint is reached.
The ambient temperature
does not meet the require-
ments.
Check the ambient temperature on the equip-
ment and adjust if necessary.
Operator
Error during temperature
measurement.
Check the function of the temperature sensor. In
case of a defect, contact your local Infors repre-
sentative.
Operator
Heater fan not
working.
Contact your local Infors representative and re-
place the control rack (see chapter 10.2, page
73).
Operator
Alarm message
Meaning
Rotation speed too low.
Possible cause
Measures
Personnel
The load is too heavy, making
the drive too weak or causing
it to overheat.
Reduce the load.
If the drive has overheated, the process can be
continued after it has cooled down.
Operator
The drive belt is torn.
Contact your local Infors representative and have
the drive belt replaced.
Infors service techni-
cian or licensed dealer
The motor is defective.
Contact your local Infors representative.
Infors service techni-
cian or licensed dealer