12.4.2 Information on Temperatures that can Actually be Reached
The temperatures that can actually be reached depend on a variety
of factors, such as the ambient temperature, the ventilation and the
temperature of the other equipment in the stack. As such, the fol-
lowing guideline values only apply to individual units (no guaran-
tees). In order to reach these guideline values, the equipment must
be free-standing and the heat it generates must be able to dissi-
pate unobstructed.
INFORMATION
The specified ambient temperature refers to the temperature di-
rectly at the equipment. In case of heat build-up due to insuffi-
cient ventilation, the temperatures on the equipment can be sig-
nificantly higher than the room temperature.
Configuration
Temperature
Individual unit with-
out cooling
Min: 5°C above ambient temperature
Max: 65°C
Individual unit with
internal cooling
Min: 10°C below ambient temperature
Max: 65°C
Individual unit with
external cooling
Depends on the operator’s
cooling cycle