Ecotron - Incubation Shaker
Safety and Responsibility
Page
18
of
90
03 May 2018
2.8 Safety Features
The equipment comes with the following safety features:
Main switch with emergency switching off
In addition to allowing you to switch the equipment on and off nor-
mally, the main switch also works as an emergency switch. If the
main switch is switched off, all circuits of the equipment are discon-
nected completely from the grid.
Equipment fuses
Two equipment fuses protect the equipment from impermissibly
high power input. The equipment fuses are located right next to the
mains connection on rear of the equipment. For descriptions of
which fuses to use for which equipment type, see chapter 10.1
"Replacing Equipment Fuses", page 72.
Door monitoring
The position of the door is monitored using a permanent magnet. If
the door is opened, all dangerous movements (shaker drive and
fan) are stopped immediately. As soon as the door is closed again
completely, the shaker drive and fan restart automatically.
Overheating shut-down
The equipment is protected against overheating by means of an
overheating shut-down. If the maximum temperature of 105°C is
exceeded, the heating is deactivated automatically.