Programming Examples
Reading Trace Data using ASCII Format (HP-IB)
Reading Trace Data using ASCII Format
(HP-IB)
This C programming example does the following:
• Open an HP-IB session at address 18.
• Clear the analyzer.
• Reset the analyzer.
• Set the input port to the 50 MHz amplitude reference.
• Set the analyzer center frequency and span to 50 MHz.
• Set the analyzer to single sweep mode.
• Trigger a sweep.
• Read the operation complete query.
• Specify units in dBm.
• Set the analyzer trace data format to ASCII format.
• Trigger a sweep.
• Read the operation complete query.
• Query the trace data using ASCII format.
• Remove the "," from the ASCII trace data for analyzing data.
• Save trace data to an ASCII file by removing the "," token.
• Close the session.
Example:
#include <stdio.h>
#include <stdlib.h>
#include <math.h>
#include <conio.h>
#include <ctype.h>
#include <string.h>
#include "visa.h"
#define hpESA_IDN_E4401B
"Hewlett-Packard, E4401B"
#define hpESA_IDN_E4411B
"Hewlett-Packard, E4411B"
Visession defaultRM, viESA;
ViStatus
errStatus;
ViChar
cIdBuff[256]={0};
char
cEnter = 0;
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...