Chapter 7
7-23
Error Messages
201 to 799: Device-Specific Errors
217
RS-232 Interface Error
An error occurred on the serial interface.
213
Span limited to 5MHz when RBW < 1kHz
In narrow (digital) resolution bandwidths, below 1 kHz,
spans greater than 5 MHz are not available.
215
TG start freq is less than 1/2 res bw
TG uncalibrated at start frequencies below 1/2 the
current resolution bandwidth.
214
TG start freq is less than 9kHz
TG uncalibrated below 9 kHz.
204
TG Frequency Limit
The tracking generator has reached the limit of its
allowable frequency range.
206
Unable to initialize flatness data
A failure occurred in setting the flatness data in the
internal EEROM. Contact your local Hewlett-Packard
sales and service office.
762
Unable to load file
A failure occurred while loading a file; the file was not
loaded.
759
Unable to load state file
A saved state file from a newer firmware revision was
attempted to be loaded into an older instrument.
752
Unable to load state from file
Loading of state from a file failed.
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...