5-78
Chapter 5
Language Reference
SENSe Subsystem
AMPLITUDE/Y Scale, Corrections, Modify, Select, Antenna
AMPLITUDE/Y Scale, Corrections, Modify, Select, Cable
AMPLITUDE/Y Scale, Corrections, Modify, Select, Other
AMPLITUDE/Y Scale, Corrections, Modify, Select, User
AMPLITUDE/Y Scale, Corrections, Modify, Edit, Point
AMPLITUDE/Y Scale, Corrections, Modify, Edit, Frequency
AMPLITUDE/Y Scale, Corrections, Modify, Edit, Amplitude
AMPLITUDE/Y Scale, Corrections, Modify, Edit, Delete Point
Merge Additional Values into the Existing Amplitude
Correction Data
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA:MERGe
<freq>,<rel_ampl>{,<freq>,<rel_ampl>}
Adds the points with the specified values to the current amplitude
correction data, allowing you to merge correction data. If too much data
is merged, as many points as possible are merged into the existing data
and then an error is reported.
• <freq> is the frequency (in Hz) where the correction should be
applied; no unit is allowed in this parameter
• <rel_ampl> is the amount of relative amplitude correction (in dB)
needed; no unit is allowed in this parameter
Remarks:
CSET number equivalents to front panel access
definitions are as follows:
CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
Delete Amplitude Correction
[:SENSe]:CORRection:CSET[1]|2|3|4:DELete
Deletes the specified correction set. If the set was On, it is turned Off.
Front Panel
Access:
AMPLITUDE/Y Scale, Corrections, Modify, Delete
Set Amplitude Correction Frequency Interpolation
[:SENSe]:CORRection:CSET[1]|2|3|4:X:SPACing
LINear|LOGarithmic
Sets the frequency interpolation to linear or logarithmic for the
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...