2-16
Chapter 2
HP ESA Status Registers
Using the Analyzer Status Registers
Figure 2-3.
1
This refers to any measurement under the
MEASURE
key.
The Operation Status Register continuously monitors the hardware
and firmware status of the analyzer, and is read-only. To query the
register, send the command
:STATus:OPERation:CONDition?
The
response will be the decimal sum of the bits which are set to 1. For
example, if bit number 9 and bit number 3 are set to 1, the decimal sum
of the 2 bits is 512 plus 8. So the decimal value 520 is returned.
The transition filter specifies which types of bit state changes in the
condition register will set corresponding bits in the event register. The
changes may be positive (from 0 to 1) or negative (from 1 to 0). Send the
Bit
Description
0
A 1 in this bit position indicates that the analyzer is
performing a self-calibration.
1,2
Unused. These bits are always set to 0.
3
A 1 in this bit position indicates that a sweep is in
progress.
4
A 1 in this bit position indicates that a measurement
1
is
in progress.
5
A 1 in this bit position indicates that a measurement
1
is
in a “wait for trigger” state.
6,7
Unused. These bits are always set to 0.
8–14
Reserved. These bits are not used by the analyzer, but are
for future use with other HP products.
15
Always Zero (0).
Bit Number
7
8
9
10
11
12
13
14
15
6 5 4 3 2 1 0
Reserved
Reserved
W
aiting
for
TRIGger
SWEeping
CALibrating
Reserved
Reserved
Reserved
Reserved
Unused Unused
Unused
Unused Unused
Reserved
Description
Always
Zero
(0)
ck766a
Operation
Condition Register
Status
STATus:OPERation:CONDition?
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...