Chapter 5
5-35
Language Reference
CALibration Subsystem
Calibrate the Tracking Generator
:CALibration:TG
:CALibration:TG?
Performs an alignment of the tracking generator assembly.
The query performs the alignment and returns a zero if the alignment
is successful.
NOTE
Before executing this command, connect a cable between rear panel
connector
AMPTD REF OUT
to the front panel
INPUT
connector for all HP
ESA spectrum analyzers except HP models E4401B and E4411B.
Except for the models indicated that do not require the cable, the
alignment will fail if the cable is not connected.
Front Panel
Access:
System, Alignments, Align Now, TG