![HP E4401B Скачать руководство пользователя страница 55](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548055.webp)
Chapter 2
2-25
HP ESA Status Registers
Using the Analyzer Status Registers
The Questionable Status Frequency Condition Register contains the
following bits:
Figure 2-12
The Questionable Status Frequency condition register continuously
monitors output frequency status of the analyzer. Condition registers
are read-only. To query the condition register, send the command
Bit
Description
0
A 1 in this bit position indicates that the synthesizer in the source
(tracking generator) is unlocked.
1
A 1 in this bit position indicates that the analyzer frequency
reference is unlocked.
2,3
Reserved. These bits are not used by the analyzer, but are for future
use with other HP products.
4
A 1 in this bit position indicates that the analyzer synthesizer is
unlocked.
5
A 1 in this bit position indicates that an invalid bandwidth setting
has been requested.
6–8
Reserved. These bits are not used by the analyzer, but are for future
use with other HP products.
9–14
Unused. These bits are always set to 0.
15
Always Zero (0).
Bit Number
7
8
9
10
11
12
13
14
15
6 5 4 3 2 1 0
Reserved
Reserved
Source
Synth
Unlocked
Reserved
Invalid
BW
Synth
Unlocked
Reserved
Reserved
Unused
Unused
Unused
Unused
Unused
Unused
Description
Always
Zero
(0)
Freq
Ref
Unlocked
ck772a
Questionable
Frequency Condition Register
Status
STATus:QUEStionable:FREQuency:CONDition?
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...