![HP E4401B Скачать руководство пользователя страница 252](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548252.webp)
5-98
Chapter 5
Language Reference
SENSe Subsystem
Time Gating Delay
[:SENSe]:SWEep:TIME:GATE:DELay <time>
[:SENSe]:SWEep:TIME:GATE:DELay?
Sets the delay time from when the gate trigger occurs to when the gate
opens. This is for EDGE triggering only.
Factory Preset
and *RST:
1
µ
s
Range:
0.3
µ
s to 429 seconds
Default Unit:
seconds
Front Panel
Access:
Sweep, Gate Setup, Edge Setup, Gate Delay
Time Gate Length
[:SENSe]:SWEep:TIME:GATE:LENGth <time>
[:SENSe]:SWEep:TIME:GATE:LENGth?
Specifies the gate time length in seconds. For EDGE triggering only.
Factory Preset
and *RST:
1
µ
s
Range:
0.3
µ
s to 429 seconds
Default Unit:
seconds
Front Panel
Access:
Sweep, Gate Setup, Edge Setup, Gate Length
Time Gate Level
[:SENSe]:SWEep:TIME:GATE:LEVel HIGH|LOW
[:SENSe]:SWEep:TIME:GATE:LEVel?
Selects the level of the gate signal. This command is for LEVel
triggering only.
Factory Preset
and *RST:
High
Front Panel
Access:
Sweep, Gate Setup, Level Setup
Time Gate Polarity
[:SENSe]:SWEep:TIME:GATE:POLarity NEGative|POSitive
[:SENSe]:SWEep:TIME:GATE:POLarity?
Selects the polarity of the gate signal. This command is for EDGE
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...