![HP E4401B Скачать руководство пользователя страница 145](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548145.webp)
Chapter 3
3-81
Programming Examples
Measuring Harmonic Distortion (RS-232)
}
void main()
{
/*Program Variables*/
ViStatus viStatus = 0;
double dFundamental = 0.0;
float fHarmV[10] ={0.0};
float fHarmDbm[10]={0.0};
float fRelAmptd[10]={0.0};
float fFundaAmptdDbm=0.0;
double dFundaAmptdV=0.0;
double dMarkerFreq = 0.0;
double dPrcntDistort =0.0;
double dSumSquare =0.0;
long lMaxHarmonic =0L;
long lNum=0L;
/*Setting default values*/
lMaxHarmonic = 5;
dFundamental = 50.0;
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...