Chapter 3
3-15
Programming Examples
Using Marker Delta Mode and Marker Minimum Search
Using Marker Delta Mode and Marker
Minimum Search
This C programming example does the following:
• Open an HP-IB session at address 18.
• Clear the analyzer.
• Reset the analyzer.
• Set the input port to the 50 MHz reference.
• Set the analyzer to single sweep mode.
• Prompt the user for the start and stop frequencies and set them.
• Trigger a sweep.
• Check for operation complete.
• Set the marker to the maximum peak.
• Set the analyzer to active delta marker mode.
• Trigger a sweep.
• Check for operation complete.
• Set the marker to minimum amplitude mode.
• Query, read and display the marker amplitude.
• Close the session.
Example:
#include <stdio.h>
#include <stdlib.h>
#include <math.h>
#include <conio.h>
#include <ctype.h>
#include <string.h>
#include "visa.h"
#define hpESA_IDN_E4401B
"Hewlett-Packard, E4401B"
#define hpESA_IDN_E4411B
"Hewlett-Packard, E4411B"
Visession defaultRM, viESA;
ViStatus
errStatus;
ViChar
cIdBuff[256] = {0};
char
cEnter = 0;
int
iResult = 0;
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...