Chapter 5
5-95
Language Reference
SENSe Subsystem
Input Port Power Gain
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Turns the internal preamp on or off.
Factory Preset
and *RST:
Off
Front Panel
Access:
AMPLITUDE/Y Scale, Int Preamp On Off
Input Port Maximum Mixer Power
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]
<ampl>|<step>|<bound>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Specifies the maximum power at the input mixer.
Factory Preset
and *RST:
−
10 dBm
Range:
–100 dBm to –10 dBm
Default Unit:
dBm
Front Panel
Access:
AMPLITUDE/Y Scale, Max Mixer Lvl
Optimize Preselector Frequency
[:SENSe]::POWer[:RF]:PADJust <freq>
[:SENSe]::POWer[:RF]:PADJust?
This command allows user-defined adjustment of the preselector
frequency to optimize its response on the signal of interest.
Factory Preset
and *RST:
0 Hz
Range:
–250 MHz to 250 MHz
Default Unit:
Hz
Remarks:
This command is available only on HP ESA models
E4404B, E4405B, E4407B, and E4408B. Use this
command for signals close to the noise level, multiple
signals close together, or for other conditions when the
preselector is not tuned to the frequency of interest.
Front Panel
Access:
AMPLITUDE/Y Scale, Presel Adjust
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...