![HP E4401B Скачать руководство пользователя страница 222](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548222.webp)
5-68
Chapter 5
Language Reference
READ Subsystem
Return Harmonic N Frequency
:READ:HARMonics:FREQuency[n]?
Returns the frequency of harmonic number n, measured in Hz.
Range:
0 Hz to the maximum frequency range of the analyzer
Default Unit:
Hz
Remarks:
The harmonic frequency precision for SCPI and the
display is four significant digits.
Front Panel
Access:
MEASURE, Harmonic Dist
Measure OBW and Transmit Frequency Error
:READ:OBWidth?
This command returns scalar results of occupied bandwidth, and
transmit frequency error.
Remarks:
The results for both values are returned in Hz and in a
comma-separated list.
Measure Occupied Bandwidth
:READ:OBWidth:OBWidth?
This command returns the value of occupied bandwidth in Hz.
Front Panel
Access:
MEASURE, Occupied BW
Measure Transmit Frequency Error
:READ:OBWidth:FERRor?
This command returns the value of transmit frequency error in Hz.
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...