![HP E4401B Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548062.webp)
2-32
Chapter 2
HP ESA Status Registers
Using the Analyzer Status Registers
Figure 2-18
The Questionable Status Integrity Condition Register continuously
monitors the calibration status of the analyzer. Condition registers are
read-only. To query the condition register, send the command
:STATus:QUEStionable:CALibration:CONDition?
The response will
be the decimal sum of the bits which are set to 1.
The transition filter specifies which types of bit state changes in the
condition register will set corresponding bits in the event register. The
changes may be positive (from 0 to 1) or negative (from 1 to 0). Send the
Bit
Description
0-2
Reserved. These bits are not used by the analyzer, but are for
future use with other HP products.
3
A 1 in this bit position indicates that the analyzer is in a state
that could lead to uncalibrated measurements. Auto coupling
may resolve this problem.
4
5–11
Reserved. These bits are not used by the analyzer, but are for
future use with other HP products.
12
A 1 in this bit position indicates that the present trace data does
not reflect the existing analyzer state. Trigger a new sweep
and/or measurement.
13, 14
Reserved. These bits are not used by the analyzer, but are for
future use with other HP products.
15
Always Zero (0).
Bit Number
7
8
9
10
11
12
13
14
15
6 5 4 3 2 1 0
IF/ADC
Over
Range
Measurement
Uncal
Reserved
Reserved
Reserved
ReservedReservedInvalid
Data
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Description
Always
Zero
(0)
Reserved
ck753a
Status Questionable Integrity Condition Register
STATus:QUEStionable:INTegrity:ENABle <num>
STATus:QUEStionable:INTegrity:ENABle?
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...