![HP E4401B Скачать руководство пользователя страница 184](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548184.webp)
5-30
Chapter 5
Language Reference
CALCulate Subsystem
Marker Read Y Value
:CALCulate:MARKer[1]|2|3|4:Y?
Read the current Y value for the designated marker or delta on its
assigned trace. The value is in the Y-axis units for the current trace
(which is often dBm).
Default Unit:
Matches the units of the trace on which the marker is
positioned
Remarks:
This command can be used to read the results of
marker functions such as band power and noise that
are displayed in the marker value field on the analyzer.
CALCulate:NTData Subsection
Normalize the Trace Data
:CALCulate:NTData[:STATe] OFF|ON|0|1
:CALCulate:NTData[:STATe]?
One sweep of trace data is copied to trace NRML, which is used as the
reference trace. Then for all subsequent trace sweeps, display trace 1
=
data collected into trace 1 – data in trace NRML.
Front Panel
Access:
View/Trace, Normalize, Normalize On Off
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...