Chapter 3
Programming Examples
Determine if an Error has Occurred
viQueryf(viESA, "*IDN?\n", "%t", &cIdBuff);
iResult = (strncmp( cIdBuff, hpESA_IDN_E4401B,
strlen(hpESA_IDN_E4401B)) && strncmp( cIdBuff,
hpESA_IDN_E4411B, strlen(hpESA_IDN_E4411B)));
if( iResult == 0 )
{
/*Set the input port to the 50 MHz amplitude reference
for the models E4411B, E4401B*/
viPrintf(viESA,"CAL:SOUR:STAT ON \n");
}
else
{
/* For the analyzers having frequency limits >= 3GHz,
prompt the user to connect the amplitude reference output to
the input*/
printf ("Connect amplitude reference output to the
input \n");
printf ("......Press any key to continue \n");
scanf( "%c",&cEnter);
/*Externally route the 50 MHz Signal*/
viPrintf(viESA,"CAL:SOUR:STAT ON \n");
}
}
/*Interrupt handler,trigger event handler */
ViStatus _VI_FUNCH sSrqHdlr(ViSession viESA, ViEventType
eventType, ViEvent ctx,ViAddr userHdlr)
{
ViUInt16 iStatusByte=0;
long lCond = 0L;
/* Make sure it is an SRQ event, ignore if stray event*/
if (eventType!=VI_EVENT_SERVICE_REQ)
{
printf ("\n Stray event type0x%1x\n",eventType);
/*Close the context before returning*/
viClose(ctx);
return VI_SUCCESS;
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...