![HP E4401B Скачать руководство пользователя страница 142](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548142.webp)
3-78
Chapter 3
Programming Examples
Measuring Harmonic Distortion (RS-232)
Measuring Harmonic Distortion (RS-232)
This C programming example does the following:
• Open a serial session at COM1/COM2
• Clear the analyzer
• Reset the analyzer.
• Set the input port to the 50 MHz amplitude reference.
• Set the analyzer center frequency to the fundamental frequency.
• Set the analyzer to 10 MHz span.
• Put the analyzer in a single sweep mode.
• Trigger a sweep.
• Check for operation complete.
• Perform the peak search
• Place the signal at the reference level using the marker-to-reference
level command and take sweep.
• Perform the thepeak search.
• Trigger a sweep.
• Check for operation complete.
• Perform the peak search.
• Perform activates signal track.
• Perform narrow span and wait.
• Check for operation complete.
• De-activate the signal track.
• Set units to dBm.
• Take a sweep and wait for the sweep completion.
• Check for operation complete.
• Perform the peak search.
• Take a sweep and wait for the sweep completion.
• Check for operation complete.
• Read the marker amplitude. This is the fundamental amplitude in
dBm.
• Change the amplitude units to volts
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...