![HP E4401B Скачать руководство пользователя страница 266](http://html.mh-extra.com/html/hp/e4401b/e4401b_programmers-manual_161548266.webp)
5-112
Chapter 5
Language Reference
STATus Subsystem
Questionable Integrity Enable
:STATus:QUEStionable:INTegrity:ENABle <number>
:STATus:QUEStionable:INTegrity:ENABle?
This command determines what bits in the Questionable Integrity
Condition Register will set bits in the Questionable Integrity Event
register, which also sets the Integrity Summary bit (bit 9) in the
Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Factory Preset
and *RST:
32767 (all 1’s)
Range:
0 to 32767
Questionable Integrity Event Query
:STATus:QUEStionable:INTegrity[:EVENt]?
This query returns the decimal value of the sum of the bits in the
Questionable Integrity Event register.
NOTE
The register requires that the equivalent PTR or NTR filters be set
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
data is cleared.
Questionable Integrity Negative Transition
:STATus:QUEStionable:INTegrity:NTRansition <number>
:STATus:QUEStionable:INTegrity:NTRansition?
This command determines what bits in the Questionable Integrity
Condition register will set the corresponding bit in the Questionable
Integrity Event register when that bit has a negative transition (1 to 0)
The variable <number> is the sum of the decimal values of the bits that
you want to enable.
Factory Preset
and *RST:
0
Range:
0 to 32767
Questionable Integrity Positive Transition
:STATus:QUEStionable:INTegrity:PTRansition <number>
:STATus:QUEStionable:INTegrity:PTRansition?
This command determines what bits in the Questionable Integrity
Condition register will set the corresponding bit in the Questionable
Integrity Event register when that bit has a negative transition (1 to 0)
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...