Chapter 5
5-55
Language Reference
INPut Subsystem
INPut Subsystem
The INPut subsystem controls the characteristics of analyzer input
ports.
Input Port Coupling
:INPut:COUPling AC|DC
:INPut:COUPling?
Selects AC or DC coupling for the front panel INPUT port. A blocking
capacitor is switched in for the AC mode.
CAUTION
Instrument damage can occur if there is a DC voltage present at the
INPUT and DC coupling is selected.
Factory Preset
and *RST:
AC
Remarks:
This command is available only on HP ESA spectrum
analyzer models E4404B or E4405B.
Front Panel
Access:
Input, Coupling AC DC
Clear the Input Overload
:INPut:PROTection:CLEar
Resets the overload protection circuitry for the input connector. There is
no query form of this command.
NOTE
This command is valid only for HP ESA models E4401B and E4411B.
The excessive input signal may have caused 15 dB of attenuation to be
switched in, or it may have completely switched the input connector out
so that it is connected to the internal reference signal.
Содержание E4401B
Страница 6: ...vi ...
Страница 14: ...xiv Contents ...
Страница 64: ...2 34 Chapter2 HP ESA Status Registers Using the Analyzer Status Registers ...
Страница 65: ...3 1 3 Programming Examples ...
Страница 118: ...3 54 Chapter3 Programming Examples Entering Amplitude Correction Data Close the session viClose viESA viClose defaultRM ...
Страница 151: ...4 1 4 Programming Command Cross References ...
Страница 286: ...5 132 Chapter5 Language Reference UNIT Subsystem ...
Страница 287: ...6 1 6 HP 8590 HP ESA Spectrum Analyzers Programming Conversion Guide ...
Страница 292: ...1 HP 8590 HP ESA Spectrum Analyzers Conversion Guide ...
Страница 368: ...7 1 7 Error Messages ...
Страница 393: ...7 26 Chapter7 Error Messages 201 to 799 Device Specific Errors ...