HP ESA Status Registers
Using the Analyzer Status Registers
NOTE
When you read the analyzer status byte register with a serial poll, the
RQS bit is reset to 0. Other bits in the register are not affected.
Restarting a measurement (
:INITiate
command) can cause the
measuring bit to pulse low, which causes an SRQ if the status register
is configured to SRQ on end-of-measurement. To avoid this:
1. Set
:INITiate:CONTinuous
off.
2. Set/enable the status registers.
3. Restart the measurement (send
:INITiate
).