Programming
Examples
2959
FOR
I=0
TO
126
2961
Exp_tbl(I+1)=Exp_tbl(I)+Exp_tbl(I)
2963
NEXT
I
2965
Exp_tbl(128)=2^(-143)
2967
FOR
I=128
TO
254
2969
Exp_tbl(I+1)=Exp_tbl(I)+Exp_tbl(I)
2971
NEXT
I
2973
RETURN
2975
Setup_fastmux:
!
2977
Setup=0
2979
ABORT
7
2981
CLEAR
7
2983
OUTPUT
@A8530_control;"RECA8"
!
START
IN
KNOWN
WORKING
STATE
2985
WAIT
3
2987
OUTPUT
@A8530_control;"HOLD;SINP"
!
GO
TO
HOLD
MODE,
SINGLE
POINT
2989
INPUT
"ENTER
FREQUENCY
OF
MEASUREMENT
(IN
GHz)",Freq
2991
OUTPUT
@A8530_control;"OUTPERRO"
!
CLEAR
THE
MESSAGE
LINE
2993
ENTER
@A8530_control;Report$
2994
!
SET
TO
SELECTED
FREQUECY,
MAKE
A
CW
MEASUREMENT:
2995
OUTPUT
@A8530_control;"CENT"&VAL$(Freq)&"GHz;SING"
2996
!
SET
FAST
MUX
DELAY
TIME
(250
uS)
2997
OUTPUT
@A8530_control;"FASPARMTIME
"&VAL$(Dwell_time)
2998
!
SET
UP
FAST
MUX
MODE
2
(SEE
O&P
MANUAL
FOR
OTHERS)
2999
OUTPUT
@A8530_control;"FASMUXMODE
2;
FASMD"
3001
OUTPUT
@A8530_control;"OUTPERRO"
!
CHECK
FOR
ERRORS
DURING
SET
UP
3003
ENTER
@A8530_control;Report$
3005
IF
VAL(Report$)<>0
THEN
3007
TABXY(1,30),"THE
FOLLOWING
PROBLEM
OCCURED
DURING
8530A
SETUP:"
3009
""
3011
Report$
3013
DISP
"RESOLVE
PROBLEM
AND
CONTINUE
(
OR
EXIT
)"
3015
FOR
N=0
TO
4
3017
ON
KEY
N
LABEL
"CONTINUE"
GOTO
Setup_fastmux
3019
ON
KEY
N+5
LABEL
"EXIT"
GOTO
Setup_failed
3021
NEXT
N
3023
Wait_for_key:GOTO
Wait_for_key
3025
END
IF
3027
REPEAT
!
WAIT
FOR
8530A
SRQ
MASK
BIT
3029
WAIT
.01
3031
UNTIL
BIT(SPOLL(716),2)
3033
TRIGGER
@A8530_control
!
DROPS
8530A
INTO
THE
SELECTED
FAST
DATA
MODE
3035
Setup=1
!
(FAST
MUX
MODE
2)
3037
Setup_failed:
!
3039
RETURN
3041
SUBEND
The
Hewlett-P
ackard
Interface
Bus
,
HP-IB
,
is
the
Hewlett-P
ackard
implementation
of
IEEE
standard
488,
dated
1978,
and
IEC
625-1.
F
or
technical
information
on
the
HP-IB
,
refer
to
the
Tutorial
Description
of
the
Hewlett-P
ackard
Interface
Bus;
P
art
Number
5952-0156.
Also
see
IEEE
standard
728-1982,
IEEE
R
ecommended
Practice
for
Code
and
F
ormat
Conventions.
18-60
HP-IB
Programming
Содержание 8530A
Страница 9: ......
Страница 31: ......
Страница 34: ...General Information Figure 1 2 Antenna Measurement Setup Using an HP 85310A GeneralInformation 1 3 ...
Страница 51: ......
Страница 65: ......
Страница 81: ......
Страница 83: ......
Страница 84: ...5 Menus Block Chapter Contents Calibration Domain Display Markers MenusBlock 5 1 ...
Страница 85: ...5 2 Menus Block ...
Страница 90: ...Calibration Requirements Figure 5 1 Cal and Cal Type Menus MenusBlock 5 7 ...
Страница 93: ...Antenna Calibration 5 10 MenusBlock ...
Страница 167: ......
Страница 195: ......
Страница 201: ......
Страница 209: ......
Страница 249: ......
Страница 281: ......
Страница 411: ......
Страница 419: ...Glossary Glossary 8 ...
Страница 439: ......