
Network
Analyzer
Calibrations
Network
Analyzer
Calibrations
This
section
explains
network
analyzer
calibrations
in
detail.
What
is
Network
Analyzer
Calibration?
Network
analyzer
calibration
greatly
reduces
repeatable
,
system-induced
errors
.
This
is
achieved
by
rst
measuring
the
magnitude
and
phase
response
of
one
or
more
high-quality
standards
.
The
standards
,
described
below
,
are
placed
(one
at
a
time)
where
the
device
under
test
(DUT)
would
normally
be
.
In
other
words
,
if
the
DUT
would
normally
be
placed
at
the
end
of
a
certain
cable
,
that's
the
exact
spot
you
should
place
the
standard
device
.
This
removes
errors
caused
by
the
cable
and
its
associated
connectors
.
This
location
is
called
the
\test
port."
The
number
of
standards
required
depends
on
the
calibration
type
.
Standards
Here
are
the
devices
that
may
be
required
during
a
network
analyzer
calibration:
Open
An
open
is
a
termination
that
presents
an
open
circuit
to
the
receiver
.
A
natural
question
to
ask
at
this
point
is:
\Why
use
a
special
termination
when
you
could
just
remove
the
device
under
test
and
leave
the
connection
open?"
The
reason
is
that
a
simple
open
connection
has
unwanted
capacitance
that
adversely
aects
the
calibration.
A
\shielded
open"
termination
solves
this
problem.
Short
A
short
is
a
termination
that
presents
a
short
circuit
at
the
test
port.
Load
There
are
two
types
of
loads
,
xed
loads
(lowband
and
broadband),
and
sliding
loads
.
Fixed
loads
are
small
terminations
that
present
a
nearly
perfect
50
ohm
impedance
.
This
absorbs
almost
all
RF
energy
that
is
produced
by
the
RF
Source
.
There
are
two
types
of
xed
load:
lowband
and
broadband.
A
sliding
load
is
a
long
device
with
a
moving
plunger
.
This
type
of
load
is
only
used
in
1-P
ort
calibrations
.
The
sliding
load
has
index
marks
along
its
length.
These
marks
allow
you
to
move
the
plunger
to
dierent
positions
required
by
the
1-P
ort
calibration.
The
device
has
a
designed-in
\mismatch"
whose
phase
vector
changes
as
you
move
the
mismatch
element
(the
mismatch
element
moves
when
you
move
the
plunger).
The
receiver
compares
the
vector
change
at
the
dierent
element
positions
,
and
interpolates
(very
accurately)
the
directivity
errors
in
the
system.
Calibration
Kits
The
calibration
standards
mentioned
above
can
be
obtained
by
purchasing
a
Hewlett-P
ackard
calibration
kit.
Cal
kits
can
only
be
used
with
the
connector
type
they
are
designed
for
.
(F
or
example
,
there
are
dierent
kits
for
3.5-mm
and
type-N
connector
types
.)
Always
use
a
cal
kit
that
has
the
proper
connector
types
for
your
test
port.
Do
not
use
adapters
with
a
cal
kit
of
the
wrong
connector
type
.
Y
our
test
port
can
have
adapters
on
it,
provided
that
the
adaptor
will
remain
in
place
when
you
attach
the
DUT
.
Data
les
provided
with
each
cal
kit
must
be
loaded
into
the
receiver
.
These
data
tell
the
receiver
the
expected
performance
values
of
the
standards
.
Cal
kit
data
les
are
also
called
\cal
kit
denitions"
in
this
manual,
because
they
dene
the
actual
performance
of
the
devices
.
After
the
receiver
measures
the
standards
,
it
compares
their
apparent
performance
with
the
information
in
the
cal
kit
denition.
Any
dierences
are
due
to
errors
in
the
system
or
cables
.
The
receiver
produces
a
list
of
the
dierences
,
called
calibration
coecients
.
When
5-24
Menus
Block
Содержание 8530A
Страница 9: ......
Страница 31: ......
Страница 34: ...General Information Figure 1 2 Antenna Measurement Setup Using an HP 85310A GeneralInformation 1 3 ...
Страница 51: ......
Страница 65: ......
Страница 81: ......
Страница 83: ......
Страница 84: ...5 Menus Block Chapter Contents Calibration Domain Display Markers MenusBlock 5 1 ...
Страница 85: ...5 2 Menus Block ...
Страница 90: ...Calibration Requirements Figure 5 1 Cal and Cal Type Menus MenusBlock 5 7 ...
Страница 93: ...Antenna Calibration 5 10 MenusBlock ...
Страница 167: ......
Страница 195: ......
Страница 201: ......
Страница 209: ......
Страница 249: ......
Страница 281: ......
Страница 411: ......
Страница 419: ...Glossary Glossary 8 ...
Страница 439: ......