Frequency/Time
Domain
Stimulus
Controls
Example:
Eects
of
Sweep
Time
If
you
use
a
far
eld
range
,
or
test
narrow-band
antennas
(such
as
narrow-band
at
plate
antennas),
determine
optimum
sweep
speed
as
follows:
1.
Set
up
the
measurement
and
notice
the
appearance
of
the
trace
at
the
Preset
sweep
time
.
2.
Store
this
trace
in
memory
by
pressing
4
DISPLA
Y
5 ,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ! MEMORY
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY:
DATA
and
MEMORY
.
3.
Then
use
the
Stimulus
menu
to
set
the
sweep
time
to
110
milliseconds/sweep
.
4.
Compare
the
new
trace
to
the
original.
Store
this
new
trace
by
pressing
4
DISPLA
Y
5 ,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DATA ! MEMORY
and
change
the
sweep
time
again.
5.
Repeat
this
process
until
you
reach
the
fastest
possible
sweep
time
with
no
change
in
the
trace
.
This
is
the
optimum
sweep
time
for
that
device
using
that
frequency
span
and
number
of
points
.
Instructions
for
storing
and
comparing
traces
are
provided
in
the
\Display"
section
of
this
manual.
Figure
6-10.
Eects
of
Sweep
Time
Stimulus
Functions
6-19
Содержание 8530A
Страница 9: ......
Страница 31: ......
Страница 34: ...General Information Figure 1 2 Antenna Measurement Setup Using an HP 85310A GeneralInformation 1 3 ...
Страница 51: ......
Страница 65: ......
Страница 81: ......
Страница 83: ......
Страница 84: ...5 Menus Block Chapter Contents Calibration Domain Display Markers MenusBlock 5 1 ...
Страница 85: ...5 2 Menus Block ...
Страница 90: ...Calibration Requirements Figure 5 1 Cal and Cal Type Menus MenusBlock 5 7 ...
Страница 93: ...Antenna Calibration 5 10 MenusBlock ...
Страница 167: ......
Страница 195: ......
Страница 201: ......
Страница 209: ......
Страница 249: ......
Страница 281: ......
Страница 411: ......
Страница 419: ...Glossary Glossary 8 ...
Страница 439: ......