Viewing
Data
Viewing
Data
from
Disc
The
following
rules
apply
to
loading
data
from
disc:
Before
loading
data
from
disc
,
place
the
receiver
in
Hold
mode
by
pressing
STIMULUS
4
MENU
5
NNNNNNNNNNNNNN
MORE
NNNNNNNNNNNNNN
HOLD
.
Otherwise
the
data
you
load
will
be
immediately
overwritten
with
new
data.
In
Frequency
Domain,
the
currently-selected
number
of
points
must
match
the
number
of
points
in
the
data
le
.
F
or
example
,
if
you
want
to
load
a
Frequency
Domain
data
le
with
801
points
,
make
sure
you
set
the
HP
8530A
to
Frequency
Domain
mode
,
and
select
STIMULUS
4
MENU
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NUMBER
of
POINTS
801
softkey
.
In
Angle
Domain,
the
current
Start,
Stop
,
and
Increment
Angle
settings
must
result
in
a
number
of
measurement
angles
that
matches
those
in
the
disc
le
.
F
or
Example
.
Y
ou
currently
have
Start
Angle
set
to
090
,
Stop
Angle
to
+90
,
and
Increment
Angle
to
1
.
This
results
in
181
measurement
angles
.
Y
ou
could
load
any
disc
le
that
has
181
angles
in
it.
T
o
load
a
le
with
a
dierent
number
of
measurement
angles
,
set
the
Start,
Stop
,
and
Increment
Angle
to
values
that
will
result
in
the
appropriate
number
of
angles
.
Note:
The
Start
Angle
,
Stop
Angle
,
and
Increment
Angle
do
not
have
to
match
those
in
the
disc
le
.
The
only
requirement
is
that
the
current
number
of
measurement
angles
must
match
the
number
of
angles
in
the
le
.
Assume
you
are
using
the
settings
listed
in
the
above
example
.
Y
ou
could
load
a
le
with
a
Start
Angle
set
to
045
,
Stop
Angle
to
+45
,
and
Increment
Angle
to
O
.5
.
In
both
cases
the
number
of
angles
in
the
measurement
is
181.
If
you
load
Raw
data,
the
receiver
places
it
in
the
Raw
data
array
,
and
performs
all
subsequent
data
processing
functions
on
it.
This
includes
calibration
(if
turned
on)
as
well
as
all
display
formatting.
After
all
processing
is
done
the
data
appears
on
the
screen.
If
you
load
\Data"
data
(corrected
data),
the
receiver
places
the
data
in
the
Corrected
data
array
,
performs
all
display
formatting.
After
all
processing
is
done
the
data
appears
on
the
screen.
If
you
want
to
load
a
display
memory
le
,
make
sure
all
memory
functions
are
o.
Y
ou
can
make
sure
of
this
by
pressing
4
DISPLA
Y
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DISPLAY:
DATA
.
This
selection
displays
current
measurement
data.
1.
Now
that
the
requirements
have
been
explained,
press:
4
DISC
5
NNNNNNNNNNNNNN
LOAD
NNNNNNNNNNNNNN
MORE
2.
Now
you
can
choose
whether
to
load
Raw
data,
\Data"
(error
corrected)
data,
or
F
ormatted
data.
Press
the
softkey
that
corresponds
the
type
of
data
you
saved
previously
.
3.
Use
the
knob
or
8
9
keys
to
select
the
le
you
wish
to
load.
4.
Press
4
LO
AD
FILE
5 .
The
data
will
appear
on
the
display
.
12-2
Viewing
Data
Содержание 8530A
Страница 9: ......
Страница 31: ......
Страница 34: ...General Information Figure 1 2 Antenna Measurement Setup Using an HP 85310A GeneralInformation 1 3 ...
Страница 51: ......
Страница 65: ......
Страница 81: ......
Страница 83: ......
Страница 84: ...5 Menus Block Chapter Contents Calibration Domain Display Markers MenusBlock 5 1 ...
Страница 85: ...5 2 Menus Block ...
Страница 90: ...Calibration Requirements Figure 5 1 Cal and Cal Type Menus MenusBlock 5 7 ...
Страница 93: ...Antenna Calibration 5 10 MenusBlock ...
Страница 167: ......
Страница 195: ......
Страница 201: ......
Страница 209: ......
Страница 249: ......
Страница 281: ......
Страница 411: ......
Страница 419: ...Glossary Glossary 8 ...
Страница 439: ......