6.2.3. Measurement
This equipment has two ways to have measurements, normal measurement and automatic
measurement. Certain numbers of measurement will carry out automatically, and you can get
the results for each measurement and their average by automatic measurement.
Normal measurement
①
Press the START key to start measurement.
②
The display will be flashing until a film reaches to the start detector.
0.000
③
Counts up the elapsed time from when the film passes the start detector.
0.123
④
After indicating the duration of the time between the film passing the start detector
and it being detected by the stop detector, the flow rate appears on the display.
1.234
Time
⇒ d−:10.23
Flow rate
⑤
Press the SELECT key to switch the indications to check the measurement data. The
following data items appear on the display every time the key is pressed.
Flow rate
⇒
Temperature
⇒
Measurement time
⇒
Atmospheric pressure
⇒
Flow rate
⇒・・・
※
Note, however, that the temperature and atmospheric pressure are displayed on
real time basis.
※
The flow rate will remain until the following measurement starts.
※
Temperatures and atmospheric pressures are sampled when the stop detector
detects the liquid films.
Automatic measurement
①
Press the AUTO key once. to enter. ("AUTO" lights up.)
②
The indicator on the AUTO key will light up. And the automatic measurement mode /
measurement count setting mode starts.
③
Use the SET keys to enter the
frequency
of automatic measurement (2 to 10 times).
AU: ※※ ※
shows the number,02
〜
10.
④
Press the START key to start the automatic measurement.
⑤
Warm-up measurement will be carried out once just after starting. Its results will be
invalid.
The display shows the number of
frequency
by”
※−
”.
(※
shows the number. And “
A
”
means 10 times.) Except for this, the display will be as same as the one on Normal
measurement.
⑥
The buzzer will sound intermittently when the measurement completed.
⑦
Press the SET key to check the measured flow rates after completing the automatic
measurement.
The display shows ”
d0
” for the average. “
d1
”
〜
”
d9
” shows 1
〜
9 times. “
dA
”
shows 10 times.
Press the SET keys to check the measured data for each measurement event.
※
To abort the automatic measurement in the midway, press the START key while
holding the SELECT key.
※
The waiting time between two consecutive measurement events is one quarter of the
preceding measurement time while the minimum and maximum are 3 and 10 seconds
respectively.
※
To reset the automatic measurement mode, press the AUTO key. The indicator on the
AUTO key will go out.
Содержание SF-1U
Страница 1: ...FILM FLOW METER SF 1U 2U Instruction Manual CODE I030248700E...
Страница 2: ...May 2016 2004 2016 HORIBA STEC Co Ltd...
Страница 3: ...J i...
Страница 4: ...J ii...
Страница 5: ...J iii 1 2 3 4 5 6 7 8 1 40 2 3 4 5 40 6 7 8 9 10 11 12 13...
Страница 7: ...J 1 SF 1U SF 2U Snoop Swagelok Company...
Страница 8: ...J 2 1 SF VP 1384428 SF 2U 2 0 35 1013 3hPa 1 3 SF 1U SF 2U 1 Snoop 59ml 1 I030248700 1 AC...
Страница 10: ...J 4 5 2...
Страница 12: ...J 6 6 6 1 6 1 1 6 1 2 AC...
Страница 13: ...J 7 6 1 3 VP 1U 4U SF 3 4 6 1 4 2 3cm 4cm Snoop Snoop...
Страница 14: ...J 8 6 1 5 VP 1U 3U 6mm 5 6mm VP 4U 10mm 9 10mm 10mm 9 10mm 3L min 2cm VP 1U 2U 1 2mm...
Страница 15: ...J 9 6 1 6 25 5 0 35 25 3 6 1 7 3 10 6 2 6 2 1 N2 5 20 1 1...
Страница 16: ...J 10 6 2 2 SF 1U 730 1070hPa 1013 3hPa SF 2U 2 730 1070hPa 1013 3hPa 2 6 2 3 2...
Страница 19: ...J 13 8 8 1...
Страница 20: ...J 14 8 2 A1 A1 A1 120 VP 1U 180 120 180 6 4 120 120 6 4 A1 7 1 3...
Страница 21: ...J 15 8 3 A2 A2 A2 120 VP 1U 180 120 180 6 4 7 1 3...
Страница 22: ...J 16 8 4 A3 A3 A3 7 1 3 8 5 A4 A4 A4...
Страница 23: ...J 17 8 6 8 7 7 1 3...
Страница 29: ...J VI B SF 1U 2U 3 3 1 RS 232C 8 1 2 NON 3 2 9600 bps 5 9600 4800 2400 1200bps 3 3 BUSY 3 4 CBM 910II 24RxxxxA...
Страница 39: ...E 4 5 Description of each part Fig 2 Description of each part...