E-5
①
Measurement gas inlet
Hose connection to measurement gas. And this is also used for pouring or
draining filming liquid.6.1.4,6.1.5
②
Measurement gas outlet
Hose connection for exhausting measurement gas.6.1.5
③
Liquid level window
For monitoring level of filming liquid.
⇒
6.1.4
④
Start detector
Film detector for starting time measurement.
⑤
Stop detector
Film detector for stopping time measurement.
⑥
Measuring tube window
For monitoring measuring tube in order to check the condition of film or dirt
inside the tube.
⑦
Film removal paper port
Port for putting in paper to remove liquid or film.6.1.5
⑧
Signal connector
(SERIAL)
For connecting printer or control signals by PC. The optional dedicated cable
is needed to connect printer or PC.
⑨
Power connector
(DC7VIN)
Connector for dedicated AC adapter.
⇒
6.1.2
⑩
Power switch
(POWER)
Switch for turning on/off.6.1.7
⑪
Fixing lever
Lever for fix/release the measuring unit.6.1.3
⑫
Display
Flow
rate
Significant digit:
4 digits
Unit selection:
999.9 ml/min or less - unit in ml/min
1000 ml/min or more - unit in ml/min
Mode display:
Normal measurement
d−:
Automatic measurement
d※:
(※
shows number of counts of automatic measurement.
2
〜
A, “A” means 10times
)
Temperature
Significant digit:
Significant digit:1
st
decimal place
Operation:
Real time display
(No hold when measurement is completed.)
Display range:
5.0
〜
40.0°C
(Temperature accuracy guaranteed 15
〜
35
℃
)
Mode display:
「1: ※※.※℃
Timing
Significant digit:
4 digits
Atmospheric
pressure
Significant digit:
Significant digit:1
st
decimal place
Operation:
Real time display
(No hold when measurement is completed.)
Display range
730 to 1070 hPa (
Range of atmospheric pressure
measurement:
930 to 1040 hPa)
Mode display:
Manual input
H.※※※※.※hPa
Atmospheric pressure measurement
A.※※※※.※hPa
(
SF-1U has only manual input function.
)
Alarm
A1:−−−−
Start detector detects abnormality.
⇒
8.2
A2:−−−−
Stop detector detects abnormality.
⇒
8.3
A3:−−−−
Detector luminous energy is low.
⇒
8.4
A4:−−−−
Voltage is low.
⇒
8.5
A1:−−−A
Automatic wetting error
⇒
8.2
Calibration
temperature
「0: ※※.※
(※
shows calibration temperature.
)⇒
6.1.6
Significant digit:
1
st
decimal place.
Range setting:
0
〜
35.0°C
Miscellaneous
Automatic
measurement
count
AU:__※※(※s
hows measurement count 02 to 10.
)
Display only when automatic measurement count is
entered.
⇒
6.2.3
Automatic
wetting operation
dP:−−−− ⇒
6.2.1
Alarm time
setting
A1: ※※※
sec
A1 alarm time setting
A2: ※※※
sec
A2 alarm time setting
(※
shows time setting.
)⇒
6.4,8.2, 8.3
⑬
Unit indicator
LED indicates the unit of value display shows.
⑭
SET key
For various settings and calling up automatic measurement data.
⑮
SELECT key
For selecting display and set items.
⑯
AUTO key
For selecting automatic measurement mode. Automatic measurement mode
is ready when LED on this key is lighted.
⇒
6.2.3
⑰
START key
Besides starting measurement/automatic measurement, this includes
functions such as alarm reset, measurement stop and automatic wetting.
Содержание SF-1U
Страница 1: ...FILM FLOW METER SF 1U 2U Instruction Manual CODE I030248700E...
Страница 2: ...May 2016 2004 2016 HORIBA STEC Co Ltd...
Страница 3: ...J i...
Страница 4: ...J ii...
Страница 5: ...J iii 1 2 3 4 5 6 7 8 1 40 2 3 4 5 40 6 7 8 9 10 11 12 13...
Страница 7: ...J 1 SF 1U SF 2U Snoop Swagelok Company...
Страница 8: ...J 2 1 SF VP 1384428 SF 2U 2 0 35 1013 3hPa 1 3 SF 1U SF 2U 1 Snoop 59ml 1 I030248700 1 AC...
Страница 10: ...J 4 5 2...
Страница 12: ...J 6 6 6 1 6 1 1 6 1 2 AC...
Страница 13: ...J 7 6 1 3 VP 1U 4U SF 3 4 6 1 4 2 3cm 4cm Snoop Snoop...
Страница 14: ...J 8 6 1 5 VP 1U 3U 6mm 5 6mm VP 4U 10mm 9 10mm 10mm 9 10mm 3L min 2cm VP 1U 2U 1 2mm...
Страница 15: ...J 9 6 1 6 25 5 0 35 25 3 6 1 7 3 10 6 2 6 2 1 N2 5 20 1 1...
Страница 16: ...J 10 6 2 2 SF 1U 730 1070hPa 1013 3hPa SF 2U 2 730 1070hPa 1013 3hPa 2 6 2 3 2...
Страница 19: ...J 13 8 8 1...
Страница 20: ...J 14 8 2 A1 A1 A1 120 VP 1U 180 120 180 6 4 120 120 6 4 A1 7 1 3...
Страница 21: ...J 15 8 3 A2 A2 A2 120 VP 1U 180 120 180 6 4 7 1 3...
Страница 22: ...J 16 8 4 A3 A3 A3 7 1 3 8 5 A4 A4 A4...
Страница 23: ...J 17 8 6 8 7 7 1 3...
Страница 29: ...J VI B SF 1U 2U 3 3 1 RS 232C 8 1 2 NON 3 2 9600 bps 5 9600 4800 2400 1200bps 3 3 BUSY 3 4 CBM 910II 24RxxxxA...
Страница 39: ...E 4 5 Description of each part Fig 2 Description of each part...