4.21 Siderail Control Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 46
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
4.21 Siderail Control Assembly
Tools required:
Phillips head screwdriver
Wire cutters
8 mm nut driver, deep well
Removal
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could cause
injury or equipment damage.
1. Disconnect the bed from its power source.
2. Disconnect the battery cable from the control board (refer to procedure 4.7
3. Remove the patient label (A) from the inner half of the siderail control
assembly (B) (see figure 4.21 on page 4-46).
Figure 4-21. Siderail Control Assembly