4.9 CPR Cable
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 20
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
4.9
CPR Cable
Tools required:
Phillips head screwdriver
4 mm hex wrench
6 mm wrench
8 mm wrench
Needle nose pliers
Wire cutters
Removal
1. Disconnect the battery cable from the control board (refer to procedure 4.7
2. Raise the bed to the highest position.
3. Raise the head section to the highest position.
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could cause
injury or equipment damage.
4. Disconnect the bed from its power source.
5. Make a note to show how the cable routes from the CPR handle assembly
(A) to the head motor (B).
CAUTION:
Use caution when you cut the wire ties. Failure to do so could cause
cable damage.
6. Carefully cut all of the wires ties that attach the CPR cable (C) to the bed
frame (D).
7. Loosen the screw (E) that attaches the CPR wire (F) to the bracket (G).
8. Remove the nut (H) that attaches the CPR cable (C) to the CPR handle
assembly (A).
9. Remove the CPR cable (C) from the CPR handle assembly (A).
10. Remove the screw (I), nut (J), and two washers (K) that attaches the CPR
wire (F) to the lever (L).