2.9 The Hilow Down Function Does Not Operate
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Page 2 - 19
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
8. Replace the controls on the siderail that has the hilow down function that
does not operate (refer to procedure 4.21 on page 4-46). The hilow down
function operates from this siderail.
Yes
No
↓
→
Replace the siderail that has the hilow down function that does
not operate (refer to procedure 4.21 on page 4-46). If this
corrects the problem, go to “Final Actions” on page 2-4.
9. Replace the control box (refer to procedure 4.1 on page 4-3). If this
corrects the problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 800-445-3720 (USA).