2.14 The CPR Function Does Not Operate
Chapter 2: Troubleshooting Procedures
2
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Page 2 - 25
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
2.14 The CPR Function Does Not Operate
1. The CPR cable (A) is connected to the CPR handle (B) (see figure 2-2 on
Yes
No
↓
→
Connect the CPR cable (A) to the CPR handle (B). If this
corrects the problem, go to “Final Actions” on page 2-4.
Otherwise, go to step 2.
2. The CPR wire (C) is held tightly in the bracket (D).
Yes
No
↓
→
Tighten the screw (E) that attaches the CPR wire (C) under the
bracket (D). If this corrects the problem, go to “Final Actions”
on page 2-4. Otherwise, go to step 3.
3. The CPR cable (A) and CPR wire (C) are connected to the head motor (F).
Yes
No
↓
→
Attach the CPR cable (A) and/or the CPR wire (C) to the head
motor (F). If this corrects the problem, go to “Final Actions” on
page 2-4. Otherwise, go to step 4.
4. Replace the head motor (F) (refer to procedure 4.8 on page 4-18). If this
corrects the problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 800-445-3720 (USA).