2.8 The Hilow Up Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Page 2 - 16
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
2.8
The Hilow Up Function Does Not Operate
1. The hilow up function does not operate when the hilow switch is pressed
from either the left siderail or the right siderail or from the foot end control
panel.
Yes
No
↓
→
2. The hilow function lockout is off.
NOTE:
If the LED for a bed function lockout is on, the function is locked out.
Yes
No
↓
→
Turn the hilow function lockout off. If this corrects the problem,
go to “Final Actions” on page 2-4.
3. Tightly seat the connector for the hilow motor into its socket on the control
box. The hilow up function does not operate.
Yes
No
↓
→
Go to “Final Actions” on page 2-4.
4. Replace the hilow motor (refer to procedure 4.11 on page 4-25). The hilow
up function does not operate.
Yes
No
↓
→
Go to “Final Actions” on page 2-4.
5. Call Hill-Rom Technical Support at 800-445-3720 (USA).
6. The hilow up function operates from the foot end control panel.
Yes
No
↓
→
Replace the foot end control panel (refer to procedure 4.2 on
page 4-5). If this corrects the problem, go to “Final Actions” on
page 2-4. Otherwise, call Hill-Rom Technical Support at
800-445-3720 (USA).
7. The hilow up function only operates from one of the siderails.
Yes
No
↓
→
Tightly seat both of the siderail connectors into their sockets on
the control box. If this corrects the problem, go to “Final
Actions” on page 2-4. Otherwise, call Hill-Rom Technical
Support at 800-445-3720 (USA).