4.18 Siderail Retraction Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 42
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
4.18 Siderail Retraction Assembly
Tools required:
4 mm hex wrench
4 mm T-handle hex wrench
Phillips head screwdriver
Screwdriver
Removal
1. Raise the bed to the highest position.
2. Lower the siderail (A) to the down position (see figure 4-18 on page 4-42).
Figure 4-18. Siderail Retraction Assembly
3. Remove the two screws (B) that attach the lower bar (C) to the siderail (A).
NOTE:
The spring assemblies will move up suddenly, and hit the top cane.
030