
6.2 CPR Cable Adjustment
Chapter 6: General Procedures
6
Hill-Rom® Basic Care™ Bed, Hill-Rom® 305 Manual Bed,
Page 6 - 11
Hill-Rom® 405 Electric Bed Service Manual (MAN336 REV 2)
6.2
CPR Cable Adjustment
Tools required:
Needle nose pliers
Small adjustable wrench
1. If the head section (A) does not lower quickly enough when you operate
the CPR handle (B), do as follows (see figure 6-2 on page 6-12):
a. Loosen the screw (C) that attaches the CPR wire (D) to the bracket (E).
b. Pull a small part of the CPR wire (D) out of the cable (F).
c. While you hold the CPR wire (D) tightly, tighten the screw (C) to
attach the wire (D) to the bracket (E).
NOTE:
While the CPR wire is held tightly, the head section will lower slowly.
d. Operate the CPR handle (B) to make sure the head section (A) lowers
at the correct rate.
e. If the head section (A) does not lower quickly enough when the CPR
handle (B) is operated, repeat step 1.
f. If the head section (A) lowers too quickly when the CPR handle (B) is
operated, go to step 2.