How to use the 5B based amplifier with Genesis and Perception
As an example of a working 5B Integration card, we will consider a card with
two 5B modules mounted, one being an f to V converter and one being a TC
module. These two modules have fixed ranges like 0
-
10.000 Hz and
-
200 °C
to +600 °C, each module has a 0
-
5 V output.
Connect the output (preferable BNCs mounted on the 5B Integration card front
panel) to two selected Genesis input channels now acting as the “Frequency”
and “TC” channels. Set sensitivity to 0
-
5 Volt (or whatever the 5B modules that
are used deliver as Full Scale value) in these channels.
Scale the inputs to 0
-
10,000 Hz and
-
200 °C to +600 °C using the TU (technical
units) input fields for those channels in Perception.
In order to do this, either calculate the TU multiplier value, or just do a two
-
point
calibration and enter the upper and lower values there.
Save this setup as a hardware setting and start any future measurements from
there.
Figure 13.6:
5B Integration card setup
CH 1A and CH 2A are scaled to measure frequency and temperature using 5B
modules in front of standard input channels.
GEN3i
I3763-3.1 en HBM: public
313
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......