B.15
Voltage probes
B.15.1
Passive, single
-
ended voltage probes
G027: Passive, Single
-
Ended Voltage Probe (Option, to be ordered separately)
To be used with single
-
ended non
-
isolated amplifiers or with differential non
-
isolated amplifiers in single
-
ended mode
10 * Uin
Uin
Passive probe
(+)
GND
GND
Single-ended amplifier
1:1
10:1
900K
100K
(+)
(+)
(-)
9M
Figure B.82:
Block diagram passive, single
-
ended voltage probe
Isolation
Not supported, always ground the negative input of the probe.
Capacitive compensation range
100 to 140 pF
DC In
-
accuracy
2%
Divide factors
1:1
10:1
Probe impedance (connected to channel)
1 MΩ
10 MΩ
-
3 dB Bandwidth
2 MHz
50 MHz
Maximum input voltage
55 V RMS CAT II
300 V RMS CAT II
Probe cable length
3 m (9.8 ft)
Probe weight
Typically 88 g (3.1 oz)
Original manufacturer's part number
PMK PMTG 323A
-
140 (LDS: 869
-
929700)
Probe operating temperature range
0 °C to +50 °C (32 °F to 122 °F)
Figure B.83:
Probe and probe accessories
GEN3i
I3763-3.1 en HBM: public
791
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......