l
Keep all measurement cables well separated from cables connected to
high switching loads or motor cables.
l
Separate measurement equipment and cables from potentially interfering
equipment like frequency inverters or wireless equipment.
General cabling remark: Only use properly rated cables to measure the signal.
Both the voltage and current rating should be matched to the signal for
measurements.
WARNING
This instrument must be properly grounded.
When using this card, we advise using the standard GEN series protective
ground connections to ensure that the entire unit is grounded.
Please see section "Connecting power" on page 71 for further details.
If connection to a protective ground is not possible for any reason, then
please refer to the international safety standard EN 50191:2000
WARNING
Overvoltage and current protection
The activation of the overvoltage protection system results in a reduced input
impedance. The overvoltage protection is inactive as long as the input voltage
is less than 200% of the selected input range.
WARNING
Disconnect voltages before removing the card from the system.
The measuring circuit can carry hazardous voltages and should be
disconnected before the card is removed from the card slot of the measurement
system.
GEN3i
242
I3763-3.1 en HBM: public
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......