Triggering
Channel trigger/qualifier
1 per channel; fully independent per channel, software selectable either trigger or qualifier
Pre
-
and post
-
trigger length
0 to full memory
Maximum trigger rate
400 triggers per second
Manual trigger (Software)
Supported
External Trigger In
Selection per card
User selectable On/Off
Active edge
Rising/Falling mainframe selectable, identical for all cards
Minimum pulse width
500 ns
Delay
±
50 ns + maximum 1 sample period
Send to External Trigger Out
User can select to forward External Trigger In to the External Trigger Out BNC
External Trigger Out
Selection per card
User selectable On/Off
Active level
High/Low/Hold High; selectable per mainframe, identical for all cards
Pulse width
High/Low: 12.8 µs
Hold high: Active from first mainframe trigger to end of recording
Pulse width created by mainframe
Delay
User selectable (3 μs to 516 µs)
±
1 µs + maximum 1 sample period (requires Perception
V6.50 or higher). The default of 516 μs is compatible with standard behavior.
Minimum selectable delay is the smallest delay available for all acquisition cards used within
the mainframe
Cross channel triggering
Measurement channels
Logical OR of triggers from all measured signals
Logical AND of qualifiers from all measured signals
Calculated channels
Logical OR of triggers from all calculated signals (RTC and RT
-
FDB)
Logical AND of qualifiers from all calculated signals (RTC and RT
-
FDB)
Analog channel trigger levels
Levels
Maximum 2 level detectors
Resolution
16 bit (0.0015%); for each level
Direction
Rising/Falling; Single direction control for both levels based on selected mode
Hysteresis
0.1 to 100% of Full Scale; defines the trigger sensitivity
Pulse detection/rejection
Disable/Detect/Reject software selectable. Maximum pulse width 65 535 samples
dY/dT conversion
dY: 16 bit (0.0015%) for both levels
dT: 1 to 1023 samples. dT setting shared for both levels
Analog channel trigger modes
Basic
POS or NEG crossing; single level
Dual level
One POS and one NEG crossing; Two individual levels, OR
-
ed
Window
Arm/trigger and a disarm level; Trigger on peak level changes in a uni
-
polar signal
Dual Window
Arm/trigger/disarm per level; Trigger on peak level changes in a bi
-
polar signal
Sequential
One arm and one trigger level; eliminate false triggering due to noise or hysteresis
Analog channel qualifier modes
Basic
Above or below level check. Enable/disable trigger with single level
Dual (level)
Outside or within bounds check. Enable/disable trigger with dual level
Trigger hold off
Disable channel trigger for 1 to 65 535 samples after trigger detected
Maximum hold off time depends on sample rate
Interval timer
Modes
Less than, trigger when rate is too low
More than, trigger when rate is too high
Between, trigger when rate between lower and upper limit
Not between, trigger when rate is not between lower and upper limit
Interval timers
Start timer and width Timer
Timer value
1 to 65 535 samples
Event counter
Counted channel trigger events before card trigger is activated
1 to 256 trigger events
GEN3i
I3763-3.1 en HBM: public
459
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......