Bridge mode and cabling
TEDS Class 3
TEDS
Class 2
optional
Bridge sensor
Shunt
-
+
R1
R2
R4
R3
Amplifier
-1
Excitation
Sense (-)
Excitation (-)
Sense (+)
Excitation (+)
120 Ω
-
¼ bridge
Signal (-)
Signal (+)
½ bridge
Signal ground
TEDS
Control
Teds (+)
Teds (-)
+
-
+V
-V
TEDS class 2/3
External shunt
CAL
Signal conditioner
+
-
+
-
U
excit
Int shunt
350
10 MΩ
10 MΩ
Ω
-
¼ bridge
Pos/neg
Int/ext
Figure 12.47:
Bridge mode block diagram
10
2
3
4
5
6
7
8
9
11
14
13
12
1
Sense
Excitation
Sense
(+)
Excitation
(+)
Signal (-)
(-)
(-)
Signal (+)
Signal ground
-
+
R1
R2
R4
R3
White
Red
Drain wires
Green
Grey
Blue
Black
Foil shield
Earth shield
12
2
11
1
7
4
8
Housing
Drain wire
Foil shield
KAB7.5/00-2/2/2
LS 2
Use double shielded wires to reduce high frequency noise
GN1640B
GN840B
Full bridge
Keep wire length < 5 cm
Figure 12.48:
Recommended 6 wire full bridge connection
GEN3i
278
I3763-3.1 en HBM: public
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......