G903: Passive, Single
-
Ended Voltage Probe (Option, to be ordered separately)
To be used with single
-
ended amplifiers or with differential amplifiers in single
-
ended mode
100 * Uin
Uin
Passive probe
(+)
GND
GND
Single-ended amplifier
900K
100K
(+)
(+)
(-)
99M
Figure B.86:
Block diagram passive, single
-
ended voltage probe
Isolation
Not supported
Capacitive compensation range
7 to 45 pF
DC In
-
accuracy
2%
Divide factors
100:1
Probe impedance (connected to channel)
100 MΩ
-
3 dB Bandwidth
400 MHz
Maximum input voltage
1000 V RMS CAT II
Probe cable length
1.2 m (3.9 ft)
Probe weight
Typically 67 g (2.4 oz)
Probe operating temperature range
0 °C to +50 °C (32 °F to 122 °F)
Original manufacturer's part number
PMK PHV1000
-
1
-
45
Figure B.87:
Probe and probe accessories
Note The compensation range of the G903 probe does not match that of GN1610/GN1611/GN3210/GN3211 cards.
For the limited bandwidth of GN1610/GN1611/GN3210/GN3211, this has no noticeable effects. When using
the G903 probe in combination with this card select the sensor "G903_NoCapacitiveCheck" from the Perception
Sensor Database.
GEN3i
I3763-3.1 en HBM: public
793
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......