Wideband (No Anti
-
Alias Protection)
When wideband is selected, there is neither an analog anti
-
alias filter nor any digital filter in the signal path. Therefore, there is no anti
-
alias
protection when wideband is selected.
Wideband bandwidth
Between 900 kHz and 1500 kHz (
-
3 dB)
0.1 dB passband flatness
(1)
DC to 160 kHz
0
Magnitude [%]
2.33
1.74
1.15
0.57
-0.57
-1.14
-1.71
-2.28
0
Magnitude [%]
2.33
1.74
1.15
0.57
-0.57
-1.14
-1.71
-2.28
Frequency [kHz]
Magnitude
[dB
]
1000
100
10
1
0.1
20 V: Wideband
passband flatness
0.2
0.1
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Magnitude
[dB
]
20 V: Wideband overview
-130
-110
-90
-70
-50
-30
-10
10
0.1
1
10
100
1000
10000
Frequency [kHz]
Magnitude
[dB
]
1000
100
10
1
0.1
4 V: Wideband
passband flatness
0.2
0.1
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Magnitude
[dB
]
4 V: Wideband overview
-130
-110
-90
-70
-50
-30
-10
10
0.1
1
10
100
1000
10000
Figure A.95:
Representative Wideband examples
(1)
Measured using a Fluke 5700A calibrator, DC normalized
GEN3i
502
I3763-3.1 en HBM: public
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......