All required accessories, can be found in the appendix “Card Options, to be
ordered separately” on page 737.
12.2.10
GN610/GN611, GN610B/GN611B Isolation and type testing
An overview of the GN610/GN611, GN610B/GN611B card isolation and input
is shown below (see Figure 12.13). The isolation of the channel to chassis is
1000 V RMS and is also qualified as 600 V CAT II (or 300 V CAT III). The
common mode of the differential input channel (isolated GND) can be
1000 V RMS with respect to the chassis. If one channel has its common mode
at +1000 V and one at
-
1000 V (with respect to chassis), the voltage between
the two channels is 2000 V. The standards at which the card is certified is
IEC61010
-
1:2010 and IEC61010
-
2
-
30:2010.
GEN3i
I3763-3.1 en HBM: public
225