Figure 10.13:
Interval timer
-
Less
A
Trigger
B
Interval
The time interval is reset on the first new trigger event. For example, this feature
allows additional pulses to be detected in a standard train of pulses.
Interval timer
-
More
This interval timer mode is more complicated. When the second trigger event
is within the set time interval, no trigger is generated and the time interval is
reset on each trigger event. When a new trigger event occurs after the specified
time interval, i.e. the interval is not reset in time, then a trigger is generated at
the end of the specified time interval.
The reset moments are denoted with a dotted line and the actual trigger moment
is denoted with a straight line.
A
B
P
Figure 10.14:
Interval timer
-
More
A
Trigger
B
Interval
GEN3i
158
I3763-3.1 en HBM: public