E.11
Unexpected mainframe shutdown
If the mainframe has shut down unexpectedly, this can have several causes:
l
Power interruption
l
Check if the mains cable is still firmly connected.
l
Check if other devices on the same mains group have experienced a
power interruption.
l
Over
-
temperature or over
-
voltage:
l
If this happens while Perception is connected, a message is shown
that tells the reason for the shutdown. Apart from that, a message is
shown when connecting the next time, even if Perception was not
connected at the time of the shutdown.
l
If the cause was over
-
temperature, please check the airflow around
the mainframe; the mainframe must be able to attract cool air to avoid
over
-
heating.
l
If the cause was over
-
voltage, please contact HBM service.
GEN3i
I3763-3.1 en HBM: public
897