Bridge mode and cabling
Shunt
Half bridge
10
2
3
4
5
6
7
8
9
11
14
13
12
1
HBM
Sense
Excitation
Sense
Excitation
(+)
(+)
(-)
(-)
(-)
Signal
Signal ground
+
-
R1
R2
Red
White
Drain wires
Green
Grey
Blue
Black
Foil shield
Earth shield
12
2
11
1
8
4
5
Housing
Drain wire
Foil shield
KAB7.5/00-2/2/2
LS 2
Use double shielded wires to reduce high frequency noise
GN1640B
GN840B
Keep wire length < 5 cm
Install shunt on LS strip
External shunt
Figure 12.49:
Recommended 6 wire half bridge with shunt resistor connection
10
2
3
4
5
6
7
8
9
11
14
13
12
1
HB
M
Quarter bridge
KAB7.5/00-2/2/2
GN1640B
GN840B
350 Ω
-
¼ bridge
Sense
Excitation
Sense
Excitation
(+)
(+)
Signal ground
-
+
R1
White
Connect
Red
Drain wires
Green
Use double shielded wires to reduce high frequency noise
Keep wire length < 5 cm
Blue
Foil shield
Earth shield
12
2
11
1
4
10
Housing
Signal (-)
(-)
(-)
8
Drain wire
LS 2
Foil shield
Figure 12.50:
Recommended 4 wire 350 Ω quarter bridge connection
GEN3i
I3763-3.1 en HBM: public
279
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......