WARNING
Measurement inputs of this instrument should not be used to measure
high
-
energy signals of measurement categories CAT II, CAT III or
CAT IV (IEC 61010
-
2
-
30:2010) (e.g. mains measurements) , unless
specifically stated for the specific input.
2.5.2
Basic insulation versus reinforced
For reference, the basic insulation and supplementary insulation and the
reinforced insulation test values for CAT II can be found below.
Table 2.2: Test voltages for the testing electric strength of solid insulation in
measuring circuits in measurement category II (IEC 61010
-
2
-
30:2010)
Nominal
voltage line to
neutral AC
RMS or DC of
MAINS being
measured. (V)
Test voltage
5 s AV test V AC RMS
Impulse test V peak
Basic
insulation and
supplementary
insulation
Reinforced
insulation
Basic
insulation and
suplementary
insulation
Reinforced
insulation
≤ 150
840
1390
1550
2500
> 150 ≤ 300
1390
2210
2500
4000
> 300 ≤ 600
2210
3510
4000
6400
> 600 ≤ 1000 3310
5400
6000
9600
Several means of protection can be used to protect a user from hazardous
voltages. As can be seen below, basic insulation and supplementary insulation
is one mean of protection, but reinforced isolation is also a means of protection.
The test voltages are different for each mean of protection, as can be found in
the table above.
GEN3i
30
I3763-3.1 en HBM: public
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......