background image

Thank you for purchasing our product.

To ensure safe and proper operation, please read this guide 

thoroughly before use.

Overview

RTR505B is a data logger designed to measure and record 

different items depending on the input module to be con

-

nected:  temperature (Thermocouple, Pt), analog signal 

(4-20mA, DC voltage), and pulse.

Recorded data is then automatically collected by the Base Unit via 

wireless communication and downloaded for archive and analysis.

RTR505B requires a Base Unit to carry out wireless communication.

 

(Compatible Base Units: RTR500BC, RTR500BW, RTR500BM, RTR-

500MBS-A, RTR-500DC, RTR-500, RTR-500NW/AW)
For the operation and configuration of the Base Unit, refer to 

the instruction manual attached to the Base Unit or the RTR500B 

Series Help available on T&D Website.

Wireless Data Recorder RTR505B is referred to as the “(data) logger” 

or “device” in this manual.

tandd.com

© Copyright T&D Corporation. All rights reserved.

 2020.11  16508210003 (1st Edition) Printed on recycled paper.

Input Modules (Sold Separately)

Measurement Item

Input Module

LCD Displayed Items

Temperature 

(Type K, J, T, S)

Thermocouple Module  

(TCM-3010)

Measurement, Unit of Measurement, 

Sensor Type, Operational Status

Temperature 

(Pt100, Pt1000)

PT Module  

(PTM-3010)

Measurement, Unit of Measurement, 

Sensor Type, Operational Status

Voltage

Voltage Module  

(VIM-3010)

Measurement, Unit of Measurement, 

Operational Status

4-20mA

4-20mA Module  

(AIM-3010)

Measurement, Unit of Measurement, 

Operational Status

Pulse

Pulse Input Cable  

(PIC-3150)

Measurement, Unit of Measurement, 

Operational Status

• Before using the pulse input cable, it is necessary to set the measurement item to “pulse 

type” in the Remote Unit Settings of the application.

• In order to change the measurement item,  initialize the Remote Unit without connecting 

the input module and then redo the Remote Unit registration and settings.

Make sure that the module is completely 

inserted until you hear a “click” sound.

Battery Installation

Recording starts automatically by inserting the battery with the 

default or previous settings.

Battery

Cover

Rubber 

Packing

Screw

Lithium Battery

Default Settings

Recording Mode

Endless

Recording Interval

10 minutes

Recording Start 

Method

Immediate Start

•  Make sure to use the proper type and 

size screwdriver. (Phillips head #1 

screwdriver is recommended.)

•  Insert the supplied battery with the 

tube attached. When using a CR2 lithi-

um battery, the tube is not necessary.

•  Before closing the cover, check the 

rubber packing for dust or scratches, 

as they can reduce the water resis-

tance of the rubber.

•  Be sure to completely close the cover. 

Make sure not to over tighten the 

screws.

•  Appropriate Tightening Torque: 

20N·cm to 30N·cm (2Kgf·cm to 3Kgf·cm)

Battery Replacement

When it is time for the battery to be 

replaced, a battery warning mark will 

appear. Please change the battery as 

soon as possible if this mark appears.

If you continue to use the logger without changing the battery,  the 

current temperature and [bAtt] will be displayed alternately and 

wireless communication will stop. (Recording will continue.)

Alternating 

Display

•  If the battery is further left unchanged, the display will automatically shut 

off and all previously recorded data will be lost.

•  Although the logger continues to work for a period after the battery has 

been removed, leaving the device without a battery until the LCD display 

goes blank will cause all recorded data to be lost.

Remote Unit Registration and Settings

Via Software and Optical Communication

Connect the Base Unit to a PC with a USB cable, and place the data 

logger face down to align the communication areas as shown below.

RTR500BW, 

RTR-500NW/AW

RTR500BM, 

RTR-500MBS-A

RTR500BC, 

RTR-500

RTR-500DC

Via Mobile App and Bluetooth

®

 Communication

When the Base Unit is a RTR500BW, it is possible to make Base 

Unit and Remote Unit settings from nearby mobile devices using 

Bluetooth.

Package Contents

• RTR505B or RTR505BL (Large Battery Type)
• Lithium Battery LS14250 (or Large Capacity Battery Kit RTR-500B1 for L type model)
• Strap (Not included with L type model)
• Manual Set (Warranty Included)

Part Names

Module Jack

LCD Display

Antenna

Optical 

Communication 

Area

<Side>

(L type)

<Side>

<Front>

<Back>

Wireless Data Recorder

RTR505B User’s Manual